| Fluid Types |
Liquids: acoustically conductive fluids, including most clean liquids, and many liquids with small amounts of entrained solids or gas bubbles |
| Flow Measurement |
Patented Correlation Transit-TimeTM model |
| Pipe Sizes |
2 to 24in (50 to 600mm) standard
Other sizes available upon request |
| Pipe Materials |
All metals and most plastics. Consult GE for concrete, composite materials, and highly corroded or lined pipes. |
| Accuracy |
±1% of reading in application
±0.5% in field calibration
Installation assumes a fully developed, symmetrical flow profile (typically 10 diameters upstream and 5 diameters downstream of straight pipe run). Final installation accuracy is a function of multiple factors including fluid, temperature range, pipe centricity, and other. |
| Calibration |
All meters are water calibrated and delivered with a traceable calibration certificate. |
|
Repeatability |
±0.2% of reading
|
| Range (Bidirectional) |
-40 to 40 ft/s (-12.19 to 12.19 m/s) |
| Rangability (Overall) |
400:1 |
| Measurement Parameters |
Velocity, Volumetric, and Totalized Flow |
| Electronics |
| Enclosure |
Epoxy-coated aluminum weatherproof Type
4X/IP67 |
| Dimensions |
6.6 x 5.0 x 2.4in (168 x 128 x 61mm)
Weight: 3.5 lbs/1.5kg |
| Channels |
One channel |
| Display |
Graphic LCD (128 x 64 pixels) |
| Keypad |
Six-button keypad, for full functionality operation |
| Error Display Indicator |
Green or red light |
| Power Supplies |
Standard: 85 to 265 VAC, 50/60 Hz
Optional: 12 to 28 VDC, ± 5%
|
| Power Consumption |
10 Watts in rush
5 Watts normal operation |
| Operating Temperature |
-4°F to 131°F (-20°C to 55°C) |
| Storage Temperature |
-40°F to 158°F (-40°C to 70°C) |
| Outputs (Based on Configuration) |
- 4-20mA (24VDC powered, 600Ω maximum load, 1500 VDC Isolation)
- Frequency, Pulse, Alarm (Passive output, 100VDC, 1A/1W maximum, 1500 VDC isolation)
- HART (FSK modulation, Category Flow, Protocol Version 7.5, Device Revision 2, MFG ID 157, Device Type Code 127, Number of device variables 34)
- Modbus/RS485 (Half-duplex, 1500 VDC Isolation)
Analog outputs Namur NE 43 compliant |
| Certification |
CE, UL, CSA, MCert |
| Clamp-On Ultrasonic Flow Transducers |
| Temperature Ranges |
Standard: -40°F to 302°F (-40°C to 150°C)
Optional: -328°F to 752°F (-200°C to 400°C)
See specific transducer for exact temperature range. |
| Mounting Fixture |
Anodized aluminum with stainless steel strapping |
| Couplant |
Solid couplant standard |
| Rating |
Standard: General purpose (IP66 or IP68)
See transducer for exact rating |
| Additional Options |
| Vitality™ PC Interface Software |
The AquaTrans AT600 communicates with a
PC via our Vitality interface program. Consult the manual for details on sites, logs, and other operations with a PC. |