RS Hydro are UK distributors of the Eutech Alpha COND 1000, this meter is capable of measuring a broad range of applications from ultrapure water to high conductivity samples. The adjustable temperature co-efficient and ultrapure water compensation option allows measurements in broad-range applications from ultrapure water to high conductivity samples enabling up to ten selectable ranges. For ease of use the pure water compensation option corrects the non-linearity of pure water temperature correction curves in ultrapure water applications. Results are given to ±1% full scale accuracy across ten different Conductivity ranges. This meter has galvanically isolated scaleable 0/4 to 20mA output for high quality on peripheral devices.
Additional features which enhance the use for the user are the choice of auto or manual temperature compensation, adjustable temperature coefficient from 0.0 to 10.0% to ensure higher accuracy and a line adjustment feature which corrects long cable resistance errors. This is an important feature as it is useful in applications involving high conductivity measurements.
The Alpha COND 1000 meter handily displays all electrode information after each successful calibration, all previous calibration data is retained in the event of unsuccessful calibrations.
This meter is covered by a 1 year warranty.
Manufactured by: Eutech Instruments
Model: Alpha COND 1000 Conductivity Controller / Transmitter
General
Use for virtually any batch or online applications where rapid, accurate control. Great for OEM/system integrator
Water Purification/Treatment
Use to treat batches on incoming process water, ultrapure water, boiler and feed water control
Industrial
Use in applications involving agriculture, chemical processing, boiler and water heaters, water-fab, microprocessor manufacturing, pharmaceuticals, pulp and paper industries and bleach manufacturing
Regulatory
Hook up to recorder for regulatory compliance data collection
Conductivity | |
Range | ... to 1.999 µS/cm ; ... to 19.99 µS/cm ; ... to 199.9 µS/cm ; ... to 1999 µS/cm ; ... to 5000 µS/cm ; ... to 19.99 mS/cm ; ... to 199.9 mS/cm |
Resolution | 0.001 µS/cm ; 0.01 µS/cm ; 0.1 µS/cm ; 1 µS/cm ; 5 µS/cm ; 0.01 mS/cm ; 0.1 mS/cm |
Accuracy | �?????????????????�????????????????�???????????????�??????????????�?????????????�????????????�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?�±1 % full scale reading |
Cell Constant | 0.01 ; 0.1 ; 1 ; 10 |
Temperature | |
Range | -9.9 to 125 °C |
Resolution | 0.1 °C |
Accuracy | ±0.5 °C |
Sensor | Pt100 / Pt1000 (jumper selectable) |
Compensation | Auto/manual (normalized at 25 °C) |
Coefficient | Ultrapure water or linear 0.00 to 10.00 % |
Set Point and Controller Functions | |
Set Point 1 (SP1) / Set Point 2 (SP2) |
... to 1.999 µS/cm or ... to 19.99 µS/cm or ... to 199.9 µS/cm or ... to 5000 µS/cm or ... to 19.99 mS/cm or ... to 199.9 mS/cm |
Switching Conductivity Hysteresis | 0 to 10 % of full scale |
Function (Switchable) | P control (pulse length/pulse frequency); limit control |
Adjustable Period With Pulse Length Controller | 0.5 to 20 sec |
Adjustable Period With Pulse Frequency Controller | 60 to 120 pulse/min |
Pickup/Dropout Delay | 0 to 2000 sec |
Contact Outputs | 3 SPDT relays |
Switching Voltage/ Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Alarm Functions | |
Function (Switchable) | Steady or fleet (pulse) |
Wash Cycle | 0.1 to 199.9 hr |
Wash Duration | 1 to 1999 sec |
Pickup Delay | 0 to 2000 sec |
Switching Voltage/ Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Electrical Data and Connections | |
Transmitter Function | 0/4 to 20 mA scalable outputs for Conductivity, galvanically isolated |
Hold Function Switch | To freeze output current and deactivate control relays |
Load | Max. 600 Ohm |
Conductivity Input | 2-pin terminal |
Connection Terminal | 5-pole, 17-pole terminal, detachable blocks |
Power Supply | |
Input | 110 VAC (jumper selectable) ; 48 to 62 Hz ; max. 7 VA (For TSCONCTP1001) 220 VAC (jumper selectable) ; 48 to 62 Hz ; max. 7 VA (For TSCONCTP1002) |
Main Fuse | Slow-blow 250 V / 100 mA |
Pollution Degree | 2 |
Transient Overvoltage Category | II |
EMC Specifications | |
Emitted Interference | According to EN 50081-1 |
Immunity to Interference | According to EN 50082-1 |
Environmental Conditions | |
Operating Temperature Range | -10 to 50 �?????????????????�????????????????�???????????????�??????????????�?????????????�????????????�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?�ºC |
Max. Releative Humidity | 80 % up to 31 °C decreasing linearly to 50 % at 40 °C |
Mechanical Specifications | |
Dimensions (WxHxD) | 96 x 96 x 175 mm |
Weight | 700 g (unit) / 800 g (packed) |
Ingress Protection | IP54 (front panel) |