The Eutech Alpha COND 2000 is the perfect piece of equipment to measure from ultra-pure water to highly conductive samples with ease. This conductivity controller features programmable temperature coefficients to give precise temperature compensation. Seven ranges can be measured from 0 to 2µS/cm to 0 to 1000 mS/cm all readings of up to three-decimal resolution are delivered at ±1% accuracy. One of the advantages to the Alpha COND 2000 is the adjustable temperature coefficients that give more precise temperature compensation as the user has the choice to select between programmable linear temperature compensation from 0.0% to 10.0%. And pure water temperature compensation to correct non-linearity of ultrapure water temperature correction curves. This controller accepts 4-cell and 2-cell electrodes: 4-cell electrode prevents the occurrence of electrode fouling in high conductivity samples and polarizing.
Other features which make the Alpha COND 2000 the ideal choice are the acceptance of 2 or 3 wire, pt100 or pt1000 RTD sensors for automatic temperature compensation and user adjustable cell constant values (k = 0.005 to k = 9.999). The meter handily displays electrode status after each calibration.
This product comes complete with a 1 year warranty.
Manufactured by: Thermo Scientific
Model: Alpha COND 2000 Conductivity Controller/Transmitter
Industrial
Use in applications involving agriculture, chemical processing, boiler and water heaters, water-fab, microprocessor manufacturing, pharmaceuticals, pulp and paper industries and bleach industries
General
Use for virtually any batch or online applications where rapid, accurate control is required. Great for OEM/system integrator
Water Purification/Treatment
Use to treat batches of incoming process water, ultrapure water, boiler and feed water control
Regulatory
Hook to recorder to document data for regulatory compliance
Conductivity | |
Range | ... to 2.000 µS/cm ; ... to 20.00 µS/cm ; ... to 200.0 µS/cm ; ... to 2000 µS/cm ; ... to 20.00 mS/cm ; ... to 200.0 mS/cm ; ... to 1000 mS/cm |
Resolution | 0.001 µS/cm ; 0.01 µS/cm ; 0.1 µS/cm ; 1 µS/cm ; 0.01 mS/cm ; 0.1 mS/cm ; 1 mS/cm |
Accuracy | ±1 % of full scale reading (±2 % >500 mS/cm) |
Temperature | |
Range | -10.0 to 125.0 °C (14.0 to 257.0 °F) |
Resolution | 0.1 °C / 0.1 °F |
Accuracy | ±0.5 °C / ±1.0 °F |
Sensor | Pt100/Pt1000 (jumper selectable); 2 or 3 wire |
Compensation | Auto/manual (normalized at 25 °C) |
Set Point and Controller Functions | |
Set Point 1 (SP1) / Set Point 2 (SP2) | ... to 2.000 µS/cm or ... to 20.00 µS/cm or ... to 200.0 µS/cm or ... to 2000 µS/cm or ... to 20.00 mS/cm or ... to 200.0 mS/cm or ... to 1000 mS/cm |
Switching Conductivity Hysteresis | 0 to 10 % of full scale |
Function (Switchable) | P/PI control (pulse length/pulse frequency/proportional integral); limit control; off |
Adjustable Period With Pulse Length Controller | 0.5 to 20 sec |
Adjustable Period With Pulse Frequency Controller | 60 to 120 pulse/min |
Integral Action Time (IAT) | 0 to 999.9 min |
Pickup/Dropout Delay | 0 to 2000 sec |
Wash Cycle | 0.1 to 200.0 hr |
Wash Duration | 1 to 2000 sec |
Contact Outputs | 1 SPDT, 3 SPST relays |
Switching Voltage/ Current/Power | 1/8 HP: at 125 VAC / max. 0.74 A / max. 93 VA 1/8 HP: at 250 VAC / max. 0.37 A / max. 93 VA |
Alarm Functions | |
Function (Switchable) | Steady or fleet (pulse) |
Pickup Delay | 0 to 2000 sec |
Switching Voltage/ Current/Power |
1/8 HP: at 125 VAC / max. 0.74 A / max. 93 VA 1/8 HP: at 250 VAC / max. 0.37 A / max. 93 VA |
Electrical Data and Connections | |
Transmitter Function | Two 0/4 to 20 mA scalable outputs for Conductivity and temperature, galvanically isolated |
CU 22 Function | 22 mA current output |
Isolated Output Voltage | 12 V ±0.5 V (max. 50 mA) |
Hold Function Switch | To freeze output current and deactivate control relays |
Load | Max. 600 Ohm |
Conductivity Input | 2-cell/4-cell terminal |
Connection Terminal | 5-pin, 8-pin, 9-pin &�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?� 13-pin terminal,�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?� detachable blocks (For TSCONCTP2000W) 5-pin, 9-pin &�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?� 19-pin terminal,�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?� detachable blocks (For TSCONCTP2000P) |
Display | |
LCD | UV coat, backlit 14 segments display with symbols for status information |
Backlight | On/off selectable with four level of brightness control |
Power Supply | |
Input | 80 to 250 VAC/DC ; 50/60 Hz ; 10 VA |
Main Fuse | 315 mA, anti-surge, 250 V |
Pollution Degree | 2 |
Transient Overvoltage Category | II |
EMC Specifications | |
Emitted Interference | According to EN 61326 |
Immunity To Interference | According to EN 61326 |
Environmental Conditions | |
Operating Temperature Range | 0 to 40 °C |
Max. Relative Humidity | 80 % up to 31 °C decreasing linearly to 50 % at 40 °C |
Mechanical Specifications | |
Dimensions (WxHxD) | 144 x 144 x 110 mm (For TSCONCTP2000W) 96 x 96 x 175 mm (For TSCONCTP2000P) |
Weight | 745 g (unit) / 1100 g (packed) (For TSCONCTP2000W) |
Ingress Protection | IP66 (NEMA 4X) (For TSCONCTP2000W) IP54 (front panel) (For TSCONCTP2000P) |