If you require an extensive measurement range of dissolved oxygen in ppm, mg/L or % saturation the Alpha DO 1000 controller/ transmitter is the perfect solution. This device is accepting of low range (0 to 10 ppm) and general range (0.5 to 40 ppm) probe, in addition this versatile controller/transmitter offers three dissolved oxygen modes: limit control. proportional pulse length control and proportional pulse frequency control.
The Alpha DO 1000 enables one point or two point DO calibration with electrode information is displayed after each successful calibration, previous calibration data is retained in the event of an unsuccessful calibration. Extensive accuracy features of the Alpha DO 1000 which enhance the users experience are automatic salinity and pressure compensation after each manual input and the option of manual of auto temperature compensation for enhanced accuracy.
Galvanically isolated, scaleable 0/4 to 20 mA output is a feature which provides high quality output on peripheral devices.
This product comes with a 1 year warranty.
Manufactured by: Eutech Instruments
Model: Alpha DO 1000 Dissolved Oxygen Controller/Transmitter
General
Useful for monitoring temperature and control of DO in wastewater treatment and aquaculture applications. Use with valves, blowers or aerators to ensure proper system operation.
Dissolved Oxygen | |
Range | 0.00 to 20.00 mg/L ; 0.0 to 200.0 % |
Resolution | 0.01 mg/L ; 0.10 % |
Accuracy | ±1.5 % of full scale reading |
Temperature | |
Range | -9.9 to 125.0 °C |
Resolution | 0.1 °C |
Accuracy | ±0.5 °C |
Sensor | Pt100 ; 3 wire |
Compensation | |
Temperature Compensation | Auto/manual |
Salinity Compensation | 0.0 to 50.0 ppt (manual setting and automatic correction) |
Pressure Compensation | kPa/mmHg (manual setting and automatic correction) |
Set Point And Controller Functions | |
Set Point 1 (SP1) / Set Point 2 (SP2) |
0.00 to 20.00 mg/L or 0.0 to 200.0 % |
Switching DO Hysteresis | 0.1 to 1.0 mg/L or 0 to 10.0 % |
Function (Switchable) | P control (pulse length/pulse frequency); limit control |
Adjustable Period With Pulse Length Controller | 0.5 to 20 sec |
Adjustable Period With Pulse Frequency Controller | 60 to 120 pulse/min |
Pickup/Dropout Delay | 0 to 2000 sec |
Contact Outputs | 3 SPDT relays |
Switching Voltage/ Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Alarm Functions | |
Function (Switchable) | Steady or fleet (pulse) |
Pickup Delay | 0 to 2000 sec |
Switching Voltage/ Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Electrical Data And Connections | |
Transmitter Function | 0/4 to 20 mA scalable outputs for Dissolved Oxygen, galvanically isolated |
Hold Function Switch | To freeze output current and deactivate control relays |
Load | Max. 600 Ohm |
DO Input | 2-pin terminal |
Connection Terminal | 5-pole, 17-pole terminal, detachable blocks |
Display | |
LCD | 7 segments display with symbols for status information |
Power Supply | |
Input | 110 VAC (jumper selectable) ; 48 to 62 Hz ; max. 7 VA (For TSDOCTP1001) 220 VAC (jumper selectable) ;48 to 62 Hz ; max. 7 VA (For TSDOCTP1002) |
Main Fuse | Slow-blow 250 V / 100 mA |
Pollution Degree | 2 |
Transient Overvoltage Category | II |
EMC Specifications | |
Emitted Interference | According to EN 50081-1 |
Immunity To Interference | According to EN 50082-1 |
Environmental Conditions | |
Operating Temperature Range | -10 to 50 °C |
Max. Relative Humidity | 80 % up to 31 °C decreasing linearly to 50 % at 40 °C |
Mechanical Specifications | |
Dimensions (WxHxD) | 96 x 96 x 175 mm |
Weight | 700 g (unit) / 800 g (packed) |