If you are looking for a resistivity controller/transmitter which offers high accuracy of ±1% across 2 ranges the Eutech Alpha RES 1000 resistivity controller/transmitter is the perfect choice. This controller/transmitter offers adjustable temperature co-efficient from 0.0 to 10.0% this ensures higher accuracy, a pure water compensation option is inbuit for added accuracy in specific applications making this device suitable for a wide range of applications. Selecting the pure water compensation option corrects non-linearity of pure water temperature correction curves in ultrapure water applications.
The Alpha RES 1000 meter clearly displays electrode information after each successful caibration and previous data is saved for the convienience of the user. This meter is handily equiped with alarm delay and individual set-point hystereses to prevent chattering, false alarms and the need for down time.
Galvanically isolated and scaleable 0/4 to 20 mA output for high quality output on peripheral devices is the perfect feature to make this controller/transmitter the perfect choice for resistivity measurements.
Manufactured by: Eutech Instruments
Model: Alpha RES 1000 Resistivity Controller/Transmitter
General
Use for virtually any batch or online applications where rapid, accurate control are required. Great for OEM/system integrator
Water Purification/Treatment
Suitable for pure, ultrapure and reagent grade water batches of incoming process water.
Regulatory
Hook to recorder to document data for regulatory compliance
Resistivity | |
Range | 0.000 to 1.999 MOhm ; 0.00 to 19.99 MOhm |
Resolution | 0.001 MOhm ; 0.01 MOhm |
Accuracy | ±1 % full scale reading |
Cell Constant | 0.1 ; 0.01 |
Temperature | |
Range | -9.9 to 125 °C |
Resolution | 0.1 °C |
Accuracy | ±0.5 °C |
Sensor | Pt100 / Pt1000 (jumper selectable) |
Compensation | Auto/manual (normalized at 25 °C) |
Coefficient | Ultrapure water or linear 0.00 to 10.00 % |
Set Point and Controller Functions | |
Set Point 1 (SP1) / Set Point 2 (SP2) | 0.000 to 1.999 MOhm or 0.00 to 19.99 MOhm |
Switching Resistivity Hysteresis | 0 to 10 % of full scale |
Function (Switchable) | P control (pulse length/pulse frequency); limit control |
Adjustable Period With Pulse Length Controller | 0.5 to 20 sec |
Adjustable Period With Pulse Frequency Controller | 60 to 120 pulse/min |
Pickup/Dropout Delay | 0 to 2000 sec |
Contact Outputs | 3 SPDT relays |
Switching Voltage/ Current/Power | Max. 250 VAC / max. 3 A / max. 600 VA |
Alarm Functions | |
Function (Switchable) | Steady or fleet (pulse) |
Wash Cycle | 0.1 to 199.9 hr |
Wash Duration | 1 to 1999 sec |
Pickup Delay | 0 to 2000 sec |
Switching Voltage/ Current/Power | Max. 250 VAC / max. 3 A / max. 600 VA |
Electrical Data And Connections | |
Transmitter Function | 0/4 to 20 mA scalable outputs for Resistivity,�?????????????�????????????�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?� galvanically isolated |
Hold Function Switch | To freeze output current and deactivate control relays |
Load | Max. 600 Ohm |
Resistivity Input | 2-pin terminal |
Connection Terminal | 5-pole, 17-pole terminal, detachable blocks |
Display | |
LCD | 7 segments display with symbols for status information |
Power Supply | |
Input | 110 VAC (jumper selectable) ;�?????????????�????????????�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?� 48 to 62 Hz ; max. 7 VA (For TSRESCTP1001) 220 VAC�?????????????�????????????�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?� (jumper selectable) ;�?????????????�????????????�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?� 48 to 62 Hz ; max. 7 VA (For TSRESCTP1002) |
Main Fuse | Slow-blow 250 V / 100 mA |
Pollution Degree | 2 |
Transient Overvoltage Category | II |
EMC Specifications | |
Emitted Interference | According to EN 50081-1 |
Immunity To Interference | According to EN 50082-1 |
Environmental Conditions | |
Operating Temperature Range | -10 to 50 �?????????????�????????????�???????????�??????????�?????????�????????�???????�??????�?????�????�???�??�?�ºC |
Max. Relative Humidity | 80 % up to 31 °C decreasing linearly to 50 % at 40 °C |
Mechanical Specifications | |
Dimensions (WxHxD) | 96 x 96 x 175 mm |
Weight | 700 g (unit) / 800 g (packed) |
Ingress Protection | IP54 (front panel) |