The Datastick Turbidity Meter has a range from 0 to 4000 NTU and a resolution of 0.1 NTU. The Wide-Range Datastick turbidity sensor provides a versatile solution for turbidity monitoring programmes. The turbidity sensor has a rugged construction, has anti-fouling characteristics and is extremely chemically resistant. A wide range of plug-and-play connection adapters are available including Modbus RTU, DeviceNet, Profibus, USB, CANopen, Ethernet. The turbidity probe can connect directly to a PLC(Programmable Logic Controller) for seamless integration with industrial control systems. It can also be connected to any computer to display data, calibrate and customise the unit without needing an intermediate analyser box. The onboard electronics are completely encapsulated and O-ring sealed for complete protection from moisture and humidity. An integral temperature sensor measures temperature independently. The turbidity sensor conforms to ISO 7027,
Manufactured by: Thermo Scientific
Model: Turbidity Sensor & Meter - Thermo Scientific
Drinking water
- Raw water monitoring
- Back wash monitoring
Wastewater effluent
Process clarity monitoring
Measurement System Performance† | |
Range | 0 to 4000 NTU |
Resolution | 0.1 NTU |
Accuracy | 1% of reading |
Step Response Time | 90% in 15 seconds |
Operational Environment | |
Temperature Range | -5 °C to 50 °C |
Maximum Pressure | 65 psig @ 50 °C |
Maximum Flow Rate | 10 ft/second |
Power Requirements‡ | |
Voltage Range | 10 to 30 VDC |
Maximum Power | 200 mW |
Typical Power | 120 mW |
Construction | |
Windows | Quartz |
Optical Topology | Detection at 90 degrees and 180 degrees to source |
O-rings | Viton® |
Sensor Head Material | Polycarbonate |
Weight | 1.2 lbs |
Units of Measure | |
Turbidity Mode | NTU |
Temperature Units | °C, °F |
Calibration†† | |
Sample | 1 and 2 point |
Zero | 1 point |
Temperature | 1 point |
Measurement Modes | |
Turbidity | 90 degree/180 degree, narrow beam, IR light |
Temperature | Automatic from -5 °C to 50 °C |
Other Configuration Options | |
Sensor Filter | 0 to 100 seconds |
Temperature Filter | 0 to 100 seconds |
Approvals and Ratings | |
Immunity & Emissions | CE Certified 89/336/EEC: CISPER 11, EN61000 (-4-2,-4-3,-4-4,-4-6, 4 8) |
Safety | cULus Listed; 367G E303570 |
Hazardous Locations | Haz Loc Class 1, Division 2, Groups A, B, C, D. T5 Max Ambient 50 °C |
† Note: Typical at 25 °C Performance unaffected by cable length ‡ Note: Class II DC power supply required †† Note: Turbidity and Temperature are precalibrated at the factory |