The oxy.IQ from Panametrics is a cost effective and reliable two-wire loop-powered transmitter. It has a 4 to 20 mA output and can measure oxygen in 10 ppm ranges and 7 percentage ranges, which can all be selected by the user. The transmitter is compact and uses proven sensor technology to accurately measure O₂ in a variety of gases, even in hazardous environments.
The oxy.IQ is controlled by a microprocessor that enables the user to select the range, perform calibration and trim outputs. The instrument can be easily programmed using the keypad and display, this keypad allows for complete functionality in hazardous areas.
The oxy.IQ’s oxygen sensor is an advanced galvanic fuel cell that provides superior performance and accuracy.
Also available as an analyser as part of a sampling system.
Manufactured by: Panametrics
Model: oxy.IQ Oxygen Transmitter
Natural gas
Semiconductor wafer machines
Coating process machines
Membrane air separators
Inert welding gases
Pure gaseous hydrocarbon streams
Process monitoring of gaseous monomers
Heat treating and bright annealing
Glove box purge and leak detection
Intrinsically safe (IS) installation | |
Intrinsically safe installations require a zener barrier, one IS cable and one non-IS cable | |
Power requirements | 24 to 28 VDC at 50 mA |
Cable | PN 704-1318: blue jacket, twisted pair with connector; 26 AWG; 2 or 10 m length |
Output | Total load must equal 250 Ω ±5% when using zener barrier |
General purpose and non-incendive (Div 2) installations | |
Standard option package or IS package, no Zener barrier or galvanic isolator is required. | |
Flamepoof installation | |
Standard option mounted in Flameproof enclosure for ATEX or IECEX locations. | |
Power Requirements | 24 to 28 VDC at 50 mA |
Cable | PN 704-1317: black jacket, twisted pair with connector; 26 AWG; 2 or 10 m length |
All installations | |
Process wetted materials | SS process unit: 316 stainless steel, Viton® O-ring, gold plated sensor electrical contacts and glass (Chemraz O-Ring Option) |
User-selectable measurement ranges |
|
All installations | |
Accuracy |
|
Repeatability |
|
Resolution | ±0.1% of range |
Linearity |
|
O2 sensor operating temperature | 32 to 113°F (0 to 45°C) |
Sample pressure | Vented to atmosphere during operation and calibration |
Atmospheric pressure effect | ±0.13% of reading per mmHg (directly proportional to absolute pressure). During calibration, pressure and flow must be kept constant. |
Process connection | 1/8 in. NPT inlet and outlet oxy.IQ Specifications |
Dimensions | 4.10 x 2.75 x 2.05 in. (104.1 x 69.9 x 52.1 mm) |
Weight | 1.35 lb (612 grams) |
Sample flow rate | 1.0 SCFH (500 cc/min) recommended for process units |
Electrical classification/certification | Intrinsically safe USA/Canada - IS for Class I, Div 1, Groups ABCD, T4 ATEX and IECEx - Ex ia IIC Ga T4 Tamb -20 to 60°C IS package, non-incendive without use of Zener barrier or galvanic isolator: - USA/Canada: Class 1, Div 2, Groups A, B, C, D; T4 Flame proof package - ATEX/IECEX Ex d IIC T4 IP66 |