AquaVent 5 Water Level Logger

  • Gauged pressure sensor for highly-accurate water level measurements 0.05% FS
  • Multiple built-in hydrophobic filters and desiccants No need to replace, reduces maintenance
  • Options for MODBUS (RS-232/RS-485) and SDI-12
  • Backwards and forwards compatible with older accessories using an adaptor
Product Options
Levelogger 5 App & App Interface
3001 DataGrabber 5
LevelSender 5
4001 SRU Solinst Readout Unit
3001 BioFoul Screen
114623

Overview

The AquaVent 5 records accurate water level and temperature measurements in shallow groundwater and surface water applications. It combines pressure and temperature sensors, hydrophobic filters and datalogger within a 22 mm x 173 mm (7/8" x 6.8") stainless steel housing with corrosion-resistant coating.

The Model 3500 AquaVent 5 uses a gauged pressure transducer; it is open to the atmosphere via a vented cable to surface. Atmospheric pressure is applied to the transducer diaphragm, providing a cancellation effect for barometric pressure. This results in actual water level recordings.

The vented cable and Solinst AquaVent 5 logger are protected from moisture by built-in desiccants and hydrophobic filters. The vented pressure transducer is made of Hastelloy®, making it extremely durable and accurate in a wide range of temperature and monitoring conditions. The sensor provides an accuracy of 0.05% FS, and can withstand 2 times over-pressure without permanent damage.

The robust memory can hold up to 150,000 data sets when programmed in Solinst Levelogger® Software.

Manufactured by: Solinst
Model: AquaVent 5 Water Level Logger

Services offered for this product

  • Sales
  • Rental
  • Installation
  • Calibration
  • Maintenance
  • IoT Cloud Compatible

Applications

Ideal for shallow applications: up to 20 m (65 ft) submergence
Aquifer characterisation: pumping tests, slug tests, etc.
Stream gauging, lake and reservoir management
Watershed, drainage basin and recharge monitoring
Storm water and runoff monitoring
Long-term water level monitoring in wells and surface water

Specification

AquaVent 5 Specifications
Level Sensor: Piezoresistive Silicon with Hastelloy Sensor
Accuracy: ± 0.05% FS Typical
Stability of Readings: Superior, low noise
Units of Measure: m, cm, ft., psi, kPa, bar, ºC, ºF
Resolution: 0.001% FS to 0.0006% FS
Normalization: Automatic Temperature Compensation
Temp. Comp. Range: 0º to 50ºC
Temperature Sensor: Platinum Resistance Temperature Detector (RTD)
Operating Temperature: -20ºC to 80ºC
Temp. Sensor Accuracy: ± 0.05ºC
Temp. Sensor Resolution: 0.003°C
Battery Life: 8 Years — based on 1 reading/minute
Clock Accuracy (typical): ± 1 minute/year (-20ºC to 80ºC)
Maximum # Readings: 150,000 sets of readings
Memory: Slate and Continuous
Communication: 57,600 bps Solinst USB, 1200 baud SDI-12, Modbus RS-485/RS-232 (various speeds)
Size: 22 mm x 173 mm (7/8" x 6.8")
Weight: 190 grams (6.7 oz.)
Corrosion Resistance: Baked-on coating using polymerization technology (inside and out)
Wetted Materials: Delrin®, Viton®, 316L stainless steel, Hastelloy, PFAS-free PTFE coating
Sampling Modes: Linear, User-Selectable with Repeat Mode, Event, Future Start, Future Stop, Real-Time View
Measurement Rates: 1/8 sec to 99 hrs
Barometric Compensation: Automatic

Accessories

Related products

FAQ's

As the Levelogger App Interface will not connect directly to the top of a Solinst datalogger, it is necessary to have connection accessories. With the Levelogger or Rainlogger, one of the following accessories is required:

  • Direct Read Cable, clients use this option to suspend dataloggers in the field when they wish to communicate downhole.
  • When working with Leveloggers that are suspended from wireline, or cord, the units must be removed from the well and connected to either a Direct Read Cable (as above) or an Optical Adaptor.
  • For the LevelVent, just connect to the unit via the Vented Cable and LevelVent Wellhead that all units are suspended from in the field – no other connection accessories are required.
  • The AquaVent communication connection is made from the App Interface to the AquaVent Wellhead using an App Connector Cable that is ordered separately.
The Levelogger App Interface is portable hardware that connects to the top of your Levelogger via a Direct Read Cable/Adaptor, to a LevelVent Wellhead, or to an AquaVent Wellhead cable. Bluetooth wireless technology allows communication between the Levelogger App Interface and the smart device with the app installed. It allows viewing/downloading of data, and programming of Solinst dataloggers, as well as many other features.
Solinst offers a wide range of dataloggers that can be used with the Leveloggger App Interface. They include the complete, current Levelogger Edge range of units (Edge, Junior Edge, LTC Edge, Rainlogger Edge), previous Gold Leveloggers/Barologgers, and the latest vented datalogger editions, the AquaVent and LevelVent.
  • Program and start dataloggers in the field.
  • Simplify data management in the field by transporting and using less equipment.
  • View real-time readings.
  • Download and e-mail logged data files.
  • Calibrate LTC Leveloggers (NEW).
  • Run the diagnostic utility (NEW) – used for troubleshooting

Leveloggers. Provides detailed information on the datalogger,
and includes options for running a self-test, doing a memory
dump and creating/emailing reports.

The most recent version of the Android App available is 1.2.5 but that is only available for download if your Android device has Android version 8 (oreo) or above. That means, any device running older Android versions can not download the most recent App-This is a restriction that Google has set.

Drivers checkup:

  1. Go to Control Panel
  2. Open Device Manager and look for the Optical Reader listed under “ports” or “other devices”. It will show up with an exclamation mark. To make sure you are looking at the right device, unplug it and plug it back in, the Optical Reader will disappear and appears from the list.
  3. Right click on the device
  4. Click on Update Driver Software (you must have Admin rights to do this, if not, ask to your IT Dept)
  5. Click on Browse My Computer
  6. The folder that contains the drivers should be highlighted in blue in the address bar. The right address should be C: Program Files x86 / Solinst / Levelogger 4_4 / Drivers. If this is not the address you see, click on Browse button and look for the Solinst folder: C: Program Files x86 / Solinst / Levelogger 4_5 / Drivers. Do not go any further, don’t select any subfolder under Drivers. Click OK.
  7. Make sure “include subfolders” check box below the address bar is checked
  8. Click Next
  9. At this time Windows should start installing the drivers from the correct folder.
  10. When finish, go back to Device Manager. If the device was previously listed under “other devices”, it will be now listed under “ports” as USB Serial Port (COM#) assigned.
  11. Please repeat the steps from 3 to 8 to make sure it was installed properly.
  12. Close Device Manager and go to Solinst Software. The USB Serial Port (COM#) assigned should be in the drop menu.
No, Solinst certifies that PFAS is not used, or knowingly present in the manufacturer, or production of their products. www.rshydro.co.uk/files/Solinst-PFAS-Statement.pdf